GE DATEX OHMEDA TEC 7 ISOFLURANE VAPORIZER
GE DATEX OHMEDA TEC 7 ISOFLURANE VAPORIZER
Fr892 (CHF)
orCall + 41 (0)27 720 41 00
Description
The GE Datex Ohmeda Tec 7 isoflurane anesthesia vaporizer delivers isoflurane efficiently and effectively. Each Tec 7 agent includes significant performance, convenience and ergonomic benefits, as well as important safety features to help improve agent delivery and reliability while reducing overall operating costs. This vaporizer provides consistent output with dial adjustment throughout the clinical flow range.
The Easy-Fil filler is designed to simplify agent filling and help minimize agent leakage during filling.
Large-diameter control dial incorporates fine graduations of 0.2% from 0 to 1% and 0.5% from 1 to 8%.
The Tec 7 vaporizer's ergonomic dial release allows left- or right-hand operation.
GE Tec 7 Isoflurane Vaporizer Specifications
Dimensions
Agent capacity
Pneumatic
Calibration: Tec 7 vaporizers for enflurane, isoflurane and halothane are calibrated at 5% v/v in 0.2% steps up to 1% and in 0.5% steps from 1% to 5% at 21°C using O2 at 5 L/min as carrier gas.
Sevoflurane calibration
The dial on the Tec 7 vaporizer for sevoflurane is calibrated at 8% v / v
The dial is marked in 0.2% steps up to 1% v / v, and in 1% steps between 1% and 8%.
Gas flow resistance
"ON: 5 L / min O2: 10 to 15 cm H2O at 21°C ± 2°C
Accuracy
5 L / min O2 at 21°C ± 2°C:
Vaporizer 5%: ± 0.25% of agent supplied or ± 15% of dial setting (whichever is greater)
Vaporizer 8%: ± 0.4% of agent supplied or ± 15% of dial setting (whichever is greater)
Flow rate: Vaporizers are designed to deliver a constant output over the entire clinical flow rate range from 200 mL / min to 15 L / min.
Operating conditions
Specifications
Condition | Used |
Stock Number | 5995 |
Two filling options | Easy-Fil and Funnel Fill. |
Height | 9.9 in (25 cm) |
Width | 4.5 in (11.4 cm) |
Depth | 8.3" (21 cm) |
Weight (dry) | 22 lb (7 kg) |
Charge | 300 mL (nominal) for loading dry vaporizers. |
Retention | 75 ml (nominal) retained by the wick system. |
"off" | insulated, no vaporizer resistance |
Temperature | 18 ° to 35 ° C / 64 ° to 95 ° F |
Humidity | 30% to 75% relative humidity (non-condensing) |