SEVOFLURANE TANK DATEX OHMEDA SEVOTEC 5
SEVOFLURANE TANK DATEX OHMEDA SEVOTEC 5
Fr892 (CHF)
orCall + 41 (0)27 720 41 00
Description
SPECIFICATION :
CALIBRATION
The vaporizers are calibrated at 21°C using an oxygen carrier gas at a flow rate of 5 liters / minute and are temperature, flow and pressure compensated within the specified operating range.
GAS FLOW RESISTANCE WITH VAPORIZER SWITCHED OFF
Resistance Vaporizer Setting Flow Temperature Temperature
10 -15 cm wg ON + 0% 5 liters / minute, oxygen 21 ± 2 ° C
LIQUID CAPACITY
The amount of liquid required to fully charge a vaporizer with dry wicks = 300 milliliters (nominal).
The amount retained by the wick system = 75 millilitres (nominal).
5% vaporizer
The amount of liquid needed to fill the vaporizer from minimum to maximum = 170 milliliters (nominal).
8% vaporizer (Sevotec)
The quantity of liquid needed to fill the vaporizer from minimum to maximum = 137 milliliters (nominal).
WEIGHT AND DIMENSIONS
ENVIRONMENTAL CONDITIONS
Operating temperature range = 18°C to 35°C
Storage temperature range = -40°C to +65°C
PERFORMANCE
Accuracy at 5 liters / minute O2 at 21°C ± 2°C:
Vaporizer 5% ± 0.25% of delivered agent or
± 15% of dial setting (whichever is greater).
Spray 8% ± 0.4% of delivered agent or
± 15% of dial setting (whichever is greater).
Specifications
Condition | Used |
Stock Number | 6741 |
Weight | 7.0 kg |
Height | 250 mm |
Depth | 164 mm |
Width | 110 mm |
Note | Spray packaging must be protected from condensation. |