VISUAL FIELD ZEISS-HUMPHREY FDT 710
VISUAL FIELD ZEISS-HUMPHREY FDT 710
Fr5,460 (CHF)
orCall + 41 (0)27 720 41 00
Description
Key features of the FDT visual field instrument include:
- Supra-threshold screening tests in less than 1 minute per eye
- Threshold tests at around 4 minutes per eye
- Easy to use; no special training required / minimal operator instruction
- No correction (trial) lens required; patients can wear their own correction or none at all
(Must only be within 6D of patient's refraction)
- No eye correction required for the opposite (untested) eye - it is automatically occluded
- Unaffected by ambient lighting - can be used in normal room lighting
- Unaffected by pupil size (as small as 2 mm)
- Extensive built-in normative age reference database
- world-class clinical validation by leading researchers in the field
- software upgradeability for future enhancements
Instrument
Dimensions: 25 cm [10 ''] wide x 48 cm [19 ''] deep x 43 cm [17 ''] high
Electrical connection: 320-IEC standard power input connector for worldwide use
Computer interface: RS-232 serial D 9-pin male connector, null-modem cable,
Operating conditions
Storage and shipping conditions
Test
Screening strategies Test:
- Supra-threshold 20° (SCREENING C-20 - 1)
- Supra-threshold 20° (SCREENING C-20 - 5)
- Supra-threshold 20° (SCREENING N-30 - 1)
- Supra-threshold 20° (SCREENING N-30 - 5)
Test threshold strategies:
- Threshold 20° (THRESHOLD C-20)
- Nasal Threshold 30 ° (THRESHOLD N-30)
- computer MONSTERS automated staircase thresholding procedure
- Staircase completion consists of at least four staircase repeats as well as upper and lower staircases
stair limits within 0.3 log10 units of each other.
- MONSTER threshold is calculated as the average of the last upper and lower presentations meeting the staircase completion criteria.
Reliability indices:
- 3 capture trials in C-20 and N-30 screening tests
- 6 capture trials in SEUIL TESTS C-20 and N-30
- False positive capture tests:
- 6 capture tests in SEUIL C-20 TESTS
- 8 capture tests in SEUIL N-30 TEST
- False negative capture tests:
- capture test contrast 0 dB (~ 100% )
- 3 capture trials in SEUIL C-20 TEST
- 5 capture trials in SEUIL N-30 TEST
Stimulus:
- 17 or 19 FDT patterns plus OD and OS capture fixation trial patterns (4 patterns per visual field quadrant plus a 5° central radius pattern)
Screening Test results:
- Field of Deviation with 4 qualitative loss classifications (within normal limits, mild relative loss, moderate relative loss, severe loss) based on age-related normative references
- Reliability indices: fixation and false-positive capture test ratios
Results Test threshold:
- Threshold (dB) Plot
- Deviation Plot with 5 probability level classifications (P> = 5% , P <5% , P <2% , P <1% , P <0.5% ) based on age-related normative references
- MD (Mean Deviation) and PSD (Pattern of Standard Deviation) global statistics Value indices with 5 probability level classifications (P> = 5% , P <5% , P <2% , P <1% , P <0.5% ) based on age-related normative references
- Reliability indices: Fixation, False positive and false negative capture test reports
Specifications
Condition | Used |
Stock Number | 5130 |
Weight | Less than 9 kg [20 lbs.] |
Patient display size | 40° horizontal from 40° square vertical |
Food | 115/230 V ~, 50/60 Hz, 0.4 / 0.2 A, |
Power cord | Approved power cord |
Printer | high-speed, high-resolution internal thermal printer |
Temperature | -20°C to +49°C, |
Humidity | 0% to 95% non-condensing |
Pressure | 700 hPa to 1060 hPa |
Altitude | Up to 3000 m above sea level |
Contrast values | P = 5% (2 times), p = 1% , maximum contrast |
Initial contrast | p = 0.5% contrast level |
Mounting and monitoring | Heijl-Krakau mounting monitor |
Capture test contrast | 6 dB (~ 50% ) |
Presentation order | Random |
Motif | 1 ° circular diameter stimulus FDT |
Contrast capture tests | 56 dB (~ 0% ) |
Pattern | 1 of 17 Random FDT patterns |
Spatial frequency | 0.25 cycles/degree, cosine modulation |
Temporal frequency | 25 Hz flicker counterphase |
Duration | 200 to 400 ms |
Color | black and white |
Average background illumination | 100 cd / m2 nominal |
Contrast range | 56 dB (~ 0% ) to 0 dB (~ 100% ) in log10 steps |
Internal interstimuli | 0 to 500 ms, Random |